翻訳と辞書
Words near each other
・ Standard person
・ Standard pistol
・ Standard Plaza
・ Standard Point Location Code
・ Standard port
・ Standard Portable Intermediate Representation
・ Standard probability space
・ Standard Project Hurricane
・ Standard Publishing
・ Standard rack
・ Standard Radio & Telefon AB
・ Standard RAID levels
・ Standard rate turn
・ Standard Recording Company
・ Standard Records
Standard Reference Method
・ Standard Register Industrial
・ Standard Reinsurance Agreement
・ Standard Romanization (Cantonese)
・ Standard RPG System
・ Standard ruler
・ Standard Sanitary Manufacturing Company
・ Standard scale
・ Standard Schaefer
・ Standard Schedules Information Manual
・ Standard Schnauzer
・ Standard School District
・ Standard score
・ Standard sea level
・ Standard Shaft, Pennsylvania


Dictionary Lists
翻訳と辞書 辞書検索 [ 開発暫定版 ]
スポンサード リンク

Standard Reference Method : ウィキペディア英語版
Standard Reference Method
The Standard Reference Method or SRM 〔"Beer 10-A Spectrophotometric Color Method", ASBC Methods of Analysis〕 is one of several systems modern brewers use to specify beer color. Determination of the SRM value involves measuring the attenuation of light of a particular wavelength (430 nm) in passing through 1 cm of the beer, expressing the attenuation as an absorption and scaling the absorption by a constant (12.7 for SRM; 25 for EBC). The SRM (or EBC) number represents a single point in the absorption spectrum of beer. As such it cannot convey full color information which would require 81 points, but it does remarkably well in this regard (it conveys 92% of spectral information) even when fruit beers are considered. Auxiliary "deviation coefficients" (see Augmented SRM below) can pick up the remainder and are necessary for fruit beers and when subtle color differences in malt beers are to be characterized.
== Measurement Method ==
The ASBC and EBC measurements are now identical (both done at the same wavelength and in the same size cuvette) but the scaling is different. A photometer or spectrophotometer is used to measure the attenuation of light at 430 nm, as it passes through 1 cm of beer contained in a standard 1 cm by 1 cm cuvette. The absorption is the log of the ratio of the intensity of the light beam entering the sample to the intensity leaving. This difference is multiplied by 12.7 in the SRM system and 25 in the EBC (see below). For example, if the light intensity leaving is one one hundredth the light intensity entering the ratio is 100, the absorption is 2 and the SRM is 25.4. The scale factor derives from the original definition of SRM discussed in the next paragraph.

The SRM number was originally, and still is, defined by "Beer color intensity on a sample free of turbidity and having the spectral characteristics of an average beer is 10 times the absorbance of the beer measured in a 1/2 inch cell with monochromatic light at 430 nanometers."() Modern spectrophotometers use 1 cm cuvettes rather than 1/2 inch ones. When a 1 cm cuvette is used, application of the Bouguer-Beer-Lambert law shows that the multiplier should be 12.7 rather than 10. When the SRM value for a beer or wort is larger than about 30 the log linear limit of some instruments using 1 cm cuvettes is approached. In such cases the sample is diluted with deionized water. Using Beer-Lambert again gives the mathematical definition of SRM in the general case as:
:SRM=12.7\times D \times A_
where D is the dilution factor (D=1 for undiluted samples, D=2 for 1:1 dilution etc.) and A_ the absorbance at 430 nm in 1 cm.
The 430-nanometer wavelength corresponds to a deep blue light, and was chosen, as was the multiplier, to make values determined in the SRM system comparable to those determined using the Lovibond system in use at the time the SRM was adopted.〔Irwin Stone, Miller, M.C. "The Standardization of Methods for the Determination of Color in Beer"ASBC Proceedings 1949〕
The SRM was adopted in 1950 by the American Society of Brewing Chemists which had recognized the need for an instrument based measurement of color unburdened by the difficulties of the Lovibond system which relies (it is still in use in many industries including brewing - malts are often labeled with the Lovibond color of laboratory worts prepared from them) on visual comparison of the sample to tinted glass discs. Beer colors measured in SRM and degrees Lovibond were, as noted above, approximately equal at the time of adoption of the SRM. Comparison of EBC (see below) and Lovibond data published by modern malsters makes it clear that the relationship between SRM and Lovibond (ºL) is:
:SRM = 1.3546 \times - 0.76.

抄文引用元・出典: フリー百科事典『 ウィキペディア(Wikipedia)
ウィキペディアで「Standard Reference Method」の詳細全文を読む



スポンサード リンク
翻訳と辞書 : 翻訳のためのインターネットリソース

Copyright(C) kotoba.ne.jp 1997-2016. All Rights Reserved.